In Ready state the system will
a. Monitor the power supplies, the safety shutter, the control panel, the footswitch connection, the
delivery fiber connection and the BRH interlock.
b. prevent the system from firing.
c. Keep the safety shutter closed.
d. Monitor the footswitch.
(If the footswitch is pressed and there are no errors, the system will go to Treatment state after verifying
that the enery-monitoring ADC is working correctly.)
4. Treatment State
The Treatment state will be entered from the Ready state when the footswitch is pressed and there are no
errors.
In Treatment state the system will do the following:
a. Stop monitoring the control panel.
b. Verify that the limits on pulse energy and pulse rate correspond to requested values.
c. If firing the Holmium laser, preheat the laser rod.
d. Open the safety shutter.
e. Set the low-energy attenuator as required.
f. Fire the laser.
g. Monitor the laser output using both energy monitoring channels.
h. Halt firing if measured pulse energy or pulse rate are outside the specified limits.