2. Standby
Standby state will be entered
• From Startup state after power up initialization is completed.
• From ready statewhen the STANDBY button is pushed.
• From READY state when the system has been idle for five minutes without a control activa-
tion.
• From error state when the error condition is cleared; for clearable errors the user must press
the STANDBY or READY button.
The system will remain in STANDBY state until the Ready button is pressed and no system fault is in effect.
The understanding of STANDBY is that the system is powered on, has passed calibration and is monitoring
safeties, but is neither primed to fire nor physically able to emit light due to mechanical and electrical
interlocks.
In Standby state the system will
• monitor power supplies, the safety shutter, the control panel, the footswitch connection, the
delivery fiber connection and the BRH interlock
• prevent the system from firing
• keep the safety shutter closed
3. Ready state
Ready state will be entered only under the following conditions:
• From Standby state when the Ready key is pressed, all safety tests have been completed and
no errors are present.
• From Treat state when treatment is ended eithout error.