EasyManua.ls Logo

LUMENIS VersaPulse PowerSuite - Check the E-Stop Button; Check the BRH Plug; Check the Blastshield; Check the Fiber Switch

LUMENIS VersaPulse PowerSuite
178 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
9HUVD3XOVH3RZHU6XLWH6HULHV6HUYLFH0DQXDO &$/$'-867$/,*1
5(9% 
3. Check the E-Stop button.
a. Depress the E-Stop button and verify the system is deactivated.
b. Turn the laser back on and verify the system goes through the self-test routine.
4. Check the BRH plug.
a. Unplug the BRH plug and verify the “REMOTE INTERLOCK” message appears on the display and
the system goes to STANDBY.
b. Depress the footswitch and verify the system is unable to fire.
c. Reconnect the BRH plug into the system and verify the “REMOTE INTERLOCK” message goes
away, then place the system into READY.
5. Check the blastshield.
a. Remove the blastshield and verify the “INSTALL DEBRIS SHIELD” message appears on the display
and the system goes to STANDBY.
b. Depress the footswitch and verify the system is unable to fire.
c. Reinsert the blastshield and verify the “INSTALL DEBRIS SHIELD” message goes away, then place
the system into READY.
6. Check the fiber switch.
a. Loosen the fiber from the fiber port and verify the “ATTACH FIBER” message appears on the dis-
play, then verify the system goes to STANDBY.
b. Depress the footswitch and verify the system is unable to fire.
c. Tighten the fiber and verify the “ATTACH FIBER” message goes away. Place the system into
READY and verify that the aiming beam is visible at the end of the fiber output. Also verify that the
aiming beam intensity properly responds to the touch screen intensity control.