At the beginning of a treatment, the moving average will be set equal to the measured
pulse energy. For each pulse thereafter, the moving average will be set equal to
0.8* previous moving average + 0.2* measured energy of the current pulse
5. System Messages
The system will advise the user if any of the following conditions exist
a. Calibration disabled.
If startup calibration was disabled, the system will display a message.
b. Service Attenuator in Beam Path.
If the service attenuator is in the beam path, the system will display a message.
c. Safety Shutter Locked Open.
If the safety shutter is locked open, the system will display a message.
d. Holmium Lamp Approaching End of Life.
If any Holmium laser rods require more than 95 Joules pump energy to deliver 2 Joules laser
energy, the system will display a message.
6. User Statistics.
The software will store in NVRAM a log of the laser use. The log will be accessible through a service
mode screen. The following items will be stored in:
• Number of each shots on each laser.
• Accumulated time with power on.
• Accumulated time lasing.