• Power up
•Self-Test
•Calibration
Once these tasks have been completed, the software will do the following:
• Display results of the diagnostics and calibration routines.
• Sound beeper, marking the completion of the power up initialization.
• Paint the user mode screen or if the service switch is set for service mode, the service screen
will be painted.
NORMAL SYSTEM OPERATION:
After the system has successfully completed the power up self-test and initialization, the system will be gin its
normal operation in either User or Service mode.
During normal operation, the system software will start performing the following on-going routines:
1. Monitor user inputs as follows:
a. Input to the control panel: identify which button is pushed and act on it.
b. Input to the remote control: identify which button is pushed and act on it.
c. Restrict operation to settings within the limits specified for the model in use.
d.Footswitch: determine if treatment is requested, or cable unattached or broken.
e. BRH interlock.
2. Input and output system data as follows:
a. Display user selected operational parameters on the control panel and the remote control if one is
present.