9HUVD3XOVH3RZHU6XLWH6HULHV6HUYLFH0DQXDO ,167$//$7,21
5(9%
Cooling air requirements
Minimum 46 cm (18 in.) from walls
Physical characteristics
Width: 46 cm (18 in.)
Length: 91 cm (36 in.)
Height: 99 cm (39 in.)
Weight: 163kg (360 lbs.)
Power cord length
8.0 m (26 ft.)
Footswitch cable length
3.7 m (12 ft.)
Environmental requirements (operating)
Maximum altitude: 3,050 m (10,000 ft.)
Temperature range: 10• C to 30• C (50• F to 86• F) - must be above dew point
Maximum humidity: 90% at 27• C (81• F) non-condensing
Environmental requirements (non-operating)
Maximum altitude: 9,144 m (34,564 ft.)
Temperature range: 10• C to 50• C (50• F to 122• F) - must be above dew point
Maximum humidity: 90% at 55• C (131• F) non-condensing
Optional external cooling system
Cooling capacity: 4.5 kW
Flow rate: 0.25 lps (4gpm)
Temperature set point: 20• C (68• F)
Connections: Quick connect
Line size: 1/2 inch minimum
Compatible delivery systems: (The laser is intended for use only with Coherent-qualified delivery sys-
tems)