Root NIBPT Chapter 3: Operation
www.masimo.com 61 Masimo
Mean Arterial Pressure (MAP)
From the Mean Arterial Pressure screen, touch Alarms, and then change any of the following
options:
Options Description
Alarm
Priority
Factory
Default
Settings
Configurable
Options
High
Limit
The High Limit is upper
threshold that triggers an
alarm.
Medium 120
28-219 in steps of 1,
or Off
When set to Off,
alarm is disabled
Low
Limit
Low Limit is the lower
threshold that triggers an
alarm.
Medium 50
27-218 in steps of 1,
or Off
When set to Off,
alarm is disabled
Intervals for NIBP
From the NIBP screen, select Intervals, then select a Set Mode: Automatic, Stat, or Schedule.
See the table below for factory default and user configurable settings:
Set Mode Description Configuration
Factory
Default
Settings
User Configurable
Settings
Automatic
Automatic interval
measurement mode will
take blood pressure
measurements once every
defined interval. See Set
Mode: Automatic on page
97.
Interval 15 min
2 min, 3 min, 4 min, 5
min, 10 min, 15 min, 30
min, 60 min, 90 min