Sample Tube Files TriStar II 3020
5-10 302-42828-01 - Dec 2012
Sample Tube Files
File > Open > Sample Tube (or click the Sample Tube tab when in Advanced format)
The Sample Tube tab provides details for setting up the default settings for a Sample Tube parameter
file.
Close button
Preview button
Save button
Refer to Common Fields and Buttons - File Menu Options, page 5-1.
Field or Button Description
Sample Tube text box Contains the description of the current file.
Replace button Refer to Common Fields and Buttons - File Menu Options, page 5-1.
Blank Tube Properties
group box
For calculated free space, enter the values for the following or use the
Load from Sample File button to import this information from an existing
sample information file:
Warm free space - sample tube gas capacity measured at room
temperature.
Cold free space - sample tube gas capacity measured with the Dewar
raised.
Non-ideality factor - factor used to compensate for the forces of
attraction between molecules in a real gas. This value is used for a
calculated free space.
Field or Button Description (continued)