Calibration TriStar II 3020
6-20 302-42828-01 - Dec 2012
Match Transducers
Unit [n] > Calibration > Match Transducers
Use to zero and match the selected transducers to the main manifold transducer. Refer to
Match Transducers, page 9-26.
Servo Valve
Unit [n] > Calibration > Servo Valve
Use to calibrate the servo valve to the manifold pressure transducer. The servo valve should always be
recalibrated after a pressure calibration has been performed. The pressure transducer should be cali-
brated before starting this calibration procedure. Refer to Servo Valve, page 9-27.