EasyManua.ls Logo

Micromeritics TriStar II 3020 - Krypton Analysis

Default Icon
439 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TriStar II 3020 Performing an Analysis
302-42828-01 - Dec 2012 3-27
8. When the analysis is complete, remove the sample tube and store (or dispose) of the sample
material as applicable.
Krypton Analysis
This applies only to units with the Krypton option installed.
Use the Krypton Analysis method when performing up to three analyses using Krypton as the adsorp-
tive gas.
1. Select Unit
[n] > Krypton Analysis.
2. The remaining instructions for Krypton Analysis are the same as Sample Analysis, page 3-25
beginning with Step 2.
Use caution when removing the sample tube if using a hanging filler rod. The sample
tube O-ring or Dewar cover may snag the filler rod retaining ring. Loosen the snag
gently; excessive force may break the tip of the filler rod.

Table of Contents

Other manuals for Micromeritics TriStar II 3020

Related product manuals