EasyManua.ls Logo

Micromeritics TriStar II 3020 - Performing a Leak Test

Default Icon
439 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TriStar II 3020 Performing a Leak Test
302-42828-01 - Dec 2012 9-15
6. Open the clamp and place it around the exhaust port flange and the exhaust filter flange. Swing
the clamp fastening screw toward the exhaust port until it fits into the slot in the other half of the
clamp. Tighten the wing nut securely by hand.
7. Turn the power to the unit on by replacing the plug back into the power source or by turning the
power switch to the ON position.
Performing a Leak Test
A service representative may request that a leak test be performed to determine if there is a system
leak. The service representative may request a copy of the report generated by this test.
The following information is displayed during the test:
Prompts on preparing the instrument for the test
Approximate time period of the test
Prompts in which an operator response is required
1. To start diagnostic testing, go to Unit [n] > Diagnostics.
2. Click the down arrow in the Test field and select System Leak Test Rev. [rev].
3. Select Report after test and Preview as the destination. Click OK.
4. Click Next. Data will be inserted into the window as they are collected.
5. A popup window indicates the test is complete. Click OK.
6. Click the Report button to display the report.