TriStar II 3020 Enable Manual Control
302-42828-01 - Dec 2012 6-11
Instrument Schematic Shortcut Menus
Each manually controlled component has a shortcut menu displaying the operations available for that
particular component such as, Open, Close, Pulse. To access the shortcut menu, hover the mouse cur-
sor over the component and right click.
Shortcut Icon Description
When the cursor changes to this icon, right click to display options for the
selected component:
For Valves 1 - 15:
Open - opens the selected valve. The valve symbol changes to green. An
alternate method is to either double click the valve or select the valve
and press the keyboard spacebar to turn it off/on.
Close - closes the selected valve. The valve symbol changes to yellow.
An alternate method is to either double click the valve or select the valve
and press the keyboard spacebar to turn it off/on.
Pulse - select to quickly turn the valve on and off allowing the operation
to proceed in small increments.
For Servo Valve:
Set - select to set the servo valve target pressure and to Dose or
Evacuate.
Open - opens the selected servo valve. The valve symbol changes to
green.
Close - closes the selected servo valve. The valve symbol changes to
solid black.
When the cursor changes to this icon, right click to display options for the
elevator:
Raise - raises the elevator. Select Raise and press the keyboard space
bar to raise/lower the elevator.
Lower - lowers the elevator. Select Lower and press the keyboard space
bar to raise/lower the elevator.
Stop - select to stop the elevator from raising or lowering.