Turning the Analyzer On and Off TriStar II 3020
2-4 302-42828-01 - Dec 2012
Side Panel
Ethernet port
RS232 connector
Turning the Analyzer On and Off
Turning the Analyzer On
1. Place the computer, monitor and printer ON/OFF switches in the ON position.
2. Place the analyzer ON/OFF switch in the ON position.
3. Turn on the pump. The pump must warm for approximately two hours before performing
analyses. For analyses that require very precise results, allow the analyzer to warm up while the
sample is being prepared.
Connector Description
Ethernet port Port for an Ethernet cable allowing communication between the analyzer
and the computer.
RS-232 connector Used by service personnel to perform calibration procedures.
Do not turn off the analyzer while initialization is in progress. Doing so may damage the
instrument.