Appendix B TriStar II 3020
B-14 302-42828-01 - Dec 2013
2556- File [file name] cannot be opened. It is currently being setup for an analysis
2556- Directory database [n] error [n]
2557- Cannot access web page.
2558- The instrument is busy. The requested operation cannot be executed.
2585- The following libraries are missing: <library-name>, <library-name>, ....
4000 Series
4002- Thermal Transpiration correction had no effect.
4003- Error Converting Pressures.
4004- Error Computing Volume Adsorbed.
Cause: The sample file is currently being analyzed and is undergoing a
critical operation.
Action: Open the sample file after the critical operation has completed.
Cause: The Micromeritics web page for DFT models cannot be accessed.
This could be caused by an ISP problem of high internet traffic.
Action: Try the operation later.
Cause: The instrument is analyzing and cannot be interrupted.
Action: Try the operation later.
Cause: This message is triggered on application start up if any of the library
files used by an application, does not exist on disk
Action: Add the library into the Libraries folder.
Cause: The Thermal transpiration correction option was selected on the
Report Options window; however, the correction did not change any
pressure by more than one percent.
Action: Deselect this option to disable this message. This correction is only
meaningful for very low pressures.
Cause: An internal processing and/or hardware error occurred during report
generation.
Action: Contact a Micromeritics service representative if this error message
continues.