TriStar II 3020 Specifying Gas Ports
302-42828-01 - Dec 2012 9-25
Specifying Gas Ports
The analyzer has gas inlets for up to three analysis gases and a dedicated inlet for helium. The gases
connected to the inlets must be specified in the analysis program. If the gases are changed on one of
the inlets, make the same change on the Unit Configuration window. The analysis program must be
kept informed of any changes in gases.
1. Select Unit
[n] > Unit Configuration.
2. In the Gas Selections group box, enter the mnemonic for the gas(es) attached to the gas inlets.
3. Click OK, then click Close.
Calibrating the System
A calibration file was created specifically for your analyzer and included with your accessories. Refer
to Reinstalling Calibration Files for an Analyzer, page 4-24. It is not necessary to recalibrate the
system unless you suspect it is out of calibration. Certain calibrations are not allowed unless performed
under the direction of a Micromeritics service representative. Those calibrations are greyed out on the
Calibration menu.
To review calibration details of the analyzer, go to Unit [n] > Unit Configuration.
The following calibrations can be performed without the assistance of a service representative:
Zero pressure
Match transducers
Servo valve