EasyManua.ls Logo

Micromeritics TriStar II 3020 - T-Plot Report Options

Default Icon
439 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Report Options Files TriStar II 3020
5-38 302-42828-01 - Dec 2012
t-Plot Report Options
The t-Plot calculation allows quantitative analysis of the area and total volume ascribed to micropores.
Matrix area (the area external to micropores) is directly determined and often proves to be a valuable
way of characterizing complex mixed materials. In the Selected Reports list box, highlight t-Plot,
then click Edit.
Field or Button Description
Thickness Curve
group box
Select the thickness curve. Click Edit to modify the values in the equation
for the selected curve. The Frenkel-Halsey-Hill thickness curve can be
applied using the Halsey option.
Reference option - select Reference and click the Edit button to define a
t-curve by entering the relative pressure and thickness values. One
predefined curve is shipped with the analysis program and is found in the
Reference directory.