Appendix B TriStar II 3020
B-10 302-42828-01 - Dec 2013
2500 Series
2500- All sample file names that can be created using the sequence number pattern
already exist. You may want to modify the next sequence number.
2501- System resources have reached a dangerously low level. Please close some
windows to avoid the loss of data.
2505- Error logger cannot be initialized. Error code [n]. Program will exit.
2506- Sample file [file name] has a 'No Analysis' status and cannot be used for this
operation
2507- The sample has an invalid status and cannot be used for degassing.
2513- Unable to read the calibration file [file name].
Cause: No more sample information files can be created using the currently
entered file name sequence number.
Action: Select Options > Sample defaults and enter another sequence
number.
Cause: A large number of windows are open and consuming the system
resources available to all applications.
Action: Close one or more windows on the screen. Contact a Micromeritics
service representative if this error message continues.
Cause: An internal processing error has occurred.
Action: Contact a Micromeritics service representative if this error message
continues.
Cause: The selected sample file does not have collected data and cannot be
used for operations, for example, reporting.
Action: Enter the name of a file with a status of Complete, Analyzing or
Entered or select a sample file from the list box.
Cause: A sample file has been selected which does not have a No Analysis or
Prepared status.
Action: Select a different file with a status of No Analysis or Prepared.
Cause: An invalid calibration file was selected or one that cannot be read.
Action: Ensure the media containing the calibration file has no problems.