Performing an Analysis TriStar II 3020
3-26 302-42828-01 - Dec 2012
QuickStart Analysis
When using the QuickStart option, it is not necessary to prepare sample files in advance. Sample files
are assigned automatically and given default conditions. File contents can be replaced if necessary
using the Browse button.
1. Select Unit [n]
> QuickStart Analysis.
2. To manually close all instrument valves, click the Close Valves button.
3. File names are assigned automatically and given default conditions. To replace the parameters
with those in another sample information file, click Browse and select the sample file.
4. Verify the information populated into the sample identification, density and mass fields. This
information is pulled from the selected sample file. The density value is applicable only if using
the Calculate method for the free space determination.
5. Edit the p
o
and Bath temperature fields, if necessary.
6. Click Report after analysis to automatically generate reports when the analysis is complete. On
the Report Settings window, select the report destination. Click OK.
7. Click Start to start the analysis. A window displays data as they are collected. A short delay is
encountered before the port status at the bottom of the screen changes from the Idle state.