EasyManua.ls Logo

Micromeritics TriStar II 3020 - Preparing for Analysis

Default Icon
439 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
TriStar II 3020 Preparing for Analysis
302-42828-01 - Dec 2012 3-17
Preparing for Analysis
The following table outlines the tasks to properly prepare for an analysis and the location of the task
procedure. It is recommended to perform the tasks in the following order:
Cleaning and Labeling Sample Tubes
Sample tubes and filler rods must be clean and dry before samples are added and weighed. The follow-
ing table indicates which materials are supplied by Micromeritics and which are supplied by the user.
The procedures following the table are recommended.
1. Preheat drying oven at 110 ºC.
2. Verify that the ultrasonic cleaning unit is clean.
3. Use 5 grams of Alconox (or other suitable detergent) per 500 mL of warm water and fill the
ultrasonic unit with enough water to cover the sample tubes and filler rods (if used). Ensure the
detergent is dissolved before placing the sample tubes and filler rods into the water. If too much
detergent is used, it may be difficult to rinse from the sample tubes.
Task Name and Location
Clean the sample tube Cleaning and Labeling Sample Tubes, page 3-17
Create the sample file Defining Sample Information Files, page 3-6
Weigh the sample Determining the Sample Mass, page 3-19
Degas the sample Degassing the Sample, page 3-21
Load sample on sample port Installing the Sample Tube, page 3-21
Fill Dewar and check LN2 level Installing the Analysis Dewar, page 3-22
Materials Supplied by Micromeritics Materials Supplied by User
Sample tube
Filler Rod (if used)
Sample tube brush
Stopper for sample tube
Sample tube rack
Sample weighing support
Sample data worksheet (copied from Appendix A
of this manual)
•Drying oven
Ultrasonic cleaning unit
•Detergent
Rubber gloves or lint-free cloth
Acetone or isopropyl alcohol
Safety glasses
Waste container
Analytical balance
Pipe cleaners

Table of Contents

Other manuals for Micromeritics TriStar II 3020

Related product manuals