TriStar II 3020 Appendix B
302-42828-01 - Dec 2013 B-11
2514- Unable to write the calibration file [file name].
2515- Warning: Changing the calibration information will affect the performance of
the instrument. Only qualified service personnel should do this. Do you wish to
proceed?
2516- Warning: Keeping a backup copy of the calibration data is recommended by
Micromeritics. Would you like to do so now?
2517- Canceling this dialog will reset the calibration state to what it was when this
dialog was first opened. Are you sure you want to cancel?
2520- No data points available for reporting.
2521- Unable to program controller.
Cause: An attempt to save calibration data has failed due to possible media
problems.
Action A: Ensure the destination location has no problems.
Action B: Choose an alternate media to save the calibration data.
Cause: The process of performing a calibration operation was started.
Action: Calibration operations should only be done by or under the direction
of qualified service personnel.
Cause: A calibration operation was performed and a backup copy is
recommended.
Action: Perform a calibration Save operation.
Cause: The calibration has not been accepted.
Action: If the calibration operation was successful, press Accept.
Cause: The selected sample file does not have collected data and cannot be
used for reporting.
Action: Select a different sample file.
Cause: A hardware malfunction has occurred.
Action: Contact a Micromeritics service representative if this error message
continues.