2522- Invalid controller application file.
2523- Programming the controller failed.
2524- CRC check failed on programming controller.
2525- Unknown error programming controller.
2526- Controller download was not successful.
2527- Controller CRC error on boot block.
2528- Controller DRAM error.
2529- Controller Com1: error.
2530- Controller Com2: error.
2531- Controller debug port error.
2532- The instrument contains a different software version. Do you want to reset it?
2533- Analyzer initialization failed.
2548- System status [n]
Cause: The application’s control file has been corrupted or deleted.
Action: Reinstall the analysis program.
Cause: An internal processing and/or hardware error has occurred.
Action: Contact a Micromeritics service representative if this error message
continues.
Cause: The application has discovered a different version of software
operating in the analyzer.
Action: If there are no analyzers other than the TriStar connected to the
computer, select
Yes and allow the updated software to load.
Cause: An internal processing and/or hardware error has occurred.
Action: Contact a Micromeritics service representative if this error message
continues.
Cause: There was a problem establishing communication with the analyzer.
Action: Ensure that the communications cable is seated firmly in the Ethernet
slot at the analyzer connection and the computer connection.
Contact a Micromeritics service representative if this error message
continues.