TriStar II 3020 Report Options Files
302-42828-01 - Dec 2012 5-39
Thickness Curve
group box (continued)
To import values from an existing thickness curve (.THK file), click Open
and select the file containing the values. The table to be imported must be
saved as ASCII text with a .THK extension. It must have a two-column
format with the relative pressures in the first column and the thickness
values in the second column. Columns must be separated by a space or a
tab.
Kruk-Jaroniec-Sayari / Halsey / Harkins and Jura / Broekhoff-de
Boer / Carbon Black STSA - select the thickness curve option and click
the Edit button. Modify the equation for the selected curve as needed.
Surface Area
group box
Select the surface area value used for thickness calculations. BET is the
most commonly used option.
Fitted thickness range
text boxes
Enter the minimum and maximum thicknesses to include in the thickness
curve. The values entered for the thicknesses can be expressed in
angstroms or nanometers. Select Options > Units to specify default units.
Refer to UNIT MENU, page 6-1.
Field or Button Description (continued)