Report Options Files TriStar II 3020
5-40 302-42828-01 - Dec 2012
Pressure Range button Click to enter minimum and maximum relative pressures.
If Use calculation assignments is not selected on the Analysis
Conditions window, all of the non-outlier points of the collected data
within the specified range are used for calculating the data for this report.
If Use calculation assignments is selected, collected data points assigned
to this report type are used.
Surface area
correction factor
text box
Enter the value to correct for surface areas that are not smooth. This brings
the values for BET surface area and micropore surface area into
accordance. For most samples, the default value of 1.000 is adequate.
t-Plot checkbox Select to have a graphical representation of data generated.
X-axis - the X-axis field shows the statistical thickness of the adsorbed
film.
Y-axis - the Y-axis field shows the quantity of gas adsorbed.
Tabular Report
checkbox
Overlay Samples
checkbox
Refer to BET/Langmuir Surface Area Report Options, page 5-32.
Cancel button
From / To text boxes
OK button
Table buttons
Refer to Common Fields and Buttons - File Menu Options, page 5-1.
Field or Button Description (continued)