TriStar II 3020 Report Options Files
302-42828-01 - Dec 2012 5-27
Apply thermal
transpiration
correction checkbox
Select to correct the temperature-induced pressure difference between the
manifold and the chilled sample tube. This option is most significant for
pressures less than approximately 1.0 mmHg. Do not use filler rods in the
sample tube when applying correction for thermal transpiration.
Always use thermal transpiration when performing micropore analyses.
Refer to CALCULATIONS, page C-1 for additional information on
thermal transpiration.
Inside diameter of
sample tube text box
Enabled when Apply thermal transpiration correction is selected. Enter
the inside diameter of the sample tube used in the analysis.
Selected Reports
list box
Select the checkbox to the left of the report names to include in the report.
Calculated pore width (w), pore radius (R) or pore diameter (D) for BJH
reports. Select Options > Units to specify default calculations.
Overlays button Refer to Multiple Sample Overlays, page 3-37.
Browse button
Close button
Edit button
Preview button
Replace button
Save button
Refer to Common Fields and Buttons - File Menu Options, page 5-1.
Field or Button Description (continued)