Appendix B TriStar II 3020
B-18 302-42828-01 - Dec 2013
4028- Dubinin calculations cannot be performed because the affinity coefficient of
the analysis gas is zero.
4029- At least two fitted data points are needed for Alpha-S calculations.
4030- Preparations failed in primary data.
4031- Not enough points with a relative pressure in the range [n,n]
4033- Not enough points to generate Dubinin Tabular Report.
4034- Fewer than 2 points available for Dubinin calculations.
Cause: Dubinin values could not be calculated because the affinity
coefficient of the analysis gas is zero.
Action: Access the Dubinin Report Adsorptive options in the sample file and
enter an appropriate value for the analysis gas.
Cause: Fewer than two data points fall within the selected Alpha-s range.
Action: Edit the calculation assignments or the fitted Alpha-s range or use a
different reference curve.
Cause: Appropriate data were not available to generate the report.
Action: This message was preceded by a different error message. Refer to the
cause/action of the preceding message.
Cause: Fewer than two data points selected for the Dubinin report fall within
the selected relative pressure range.
Action: Edit the calculation assignments or the fitted relative pressure range.
Cause: There are fewer than two valid data points available for the Dubinin
tabular reports.
Action: Examine the calculation assignments in the Collected Data window
of the sample file. At least two micropore pressures must be selected
for inclusion in the Dubinin report.
Cause: There are fewer than two valid data points available for Dubinin
reports in one of the sample files selected for overlaying.
Action: Examine the calculation assignments in the Collected Data window
of the sample file. At least two micropore pressures must be selected
for inclusion in the Dubinin report.