Appendix B TriStar II 3020
B-22 302-42828-01 - Dec 2013
4048- Warning: An error occurred while restoring the heat of adsorption report editor.
4049- The sample [file name] does not have enough data. A minimum of two adsorp-
tion points is required.
4050- None of the requested quantities adsorbed is within the range of the collected
data of more than one sample file.
4051- The sample [file name] does not have any data in the range of the requested
quantities adsorbed.
4052- Fewer than two points are selected for this report.
4053- At least two data points must be selected for t-Plot calculations.
Cause: The state of the heat of adsorption report editor could not be restored.
Default settings will be used.
Action: No action.
Cause: A sample file has been included in the Heat of Adsorption report that
does not have enough data.
Action: Remove the file from the selected file list.
Cause: The Heat of Adsorption report failed because the specified quantities
are not within the range of the collected data.
Action: Edit the quantities adsorbed so that they are within the range of the
collected data or select other sample files.
Cause: The sample’s data cannot be interpolated to any of the quantities
adsorbed.
Action: Edit the quantities adsorbed so that they are within the range of the
collected data.
Cause: At least two points are required for the BET calculations.
Action: Edit the calculation assignments for the BET report.
Cause: At least two points are required for the t-Plot calculations.
Action: Edit the calculation assignments for the t-Plot report.