TriStar II 3020 Appendix B
302-42828-01 - Dec 2013 B-29
6010- This sample requires a different adsorptive and cannot be analyzed at the same
time as the other samples.
6011- The adsorptive required by this analysis is not available on this instrument.
6012- Cannot read the analysis conditions parameter file.
6013- Cannot read the adsorptive properties parameter file.
6014- Cannot read the report options parameter file.
6015- Cannot read the sample tube properties parameter file.
Cause: A sample file for analysis was selected that requires a different
adsorptive gas than the sample files selected for the other ports.
Action: Select only sample files to be analyzed with the same adsorptive and
perform the analysis. Then perform the analysis with the other
adsorptive.
Cause: An attempt was made to start an analysis with an adsorptive that is
not connected to the instrument or has not been designated in the
software.
Action A: Ensure that the adsorptive is connected to the instrument. Then select
Unit [n] > Unit configuration to tell the application that the gas is
connected.
Action B: Select only sample files for which the analysis gas is available.
Cause: The parameter file is either corrupt or has been deleted.
Action A: If this is a file created in your lab, recreate the file.
Action B: If this is a default file created during application installation, re-install
the software.
Cause: The selected sample tube file on the QuickStart screen cannot be read.
Action: Select a different file.