TriStar II 3020 Appendix B
302-42828-01 - Dec 2013 B-31
6021- Servo calibration failed.
6022- This file already selected for the analysis.
6023- The instrument [number] is not calibrated.
6024- Evacuation failed.
Cause A: The maximum time was exceeded before the target pressure point
was reached. The nitrogen regulator may be set too low or turned off.
Action A: Set the analysis gas regulator to at least 10 psig (0.7 bar), then resume
the analysis.
Cause B: The analysis gas bottle is empty.
Action B: Connect a new analysis gas bottle, then resume the analysis.
Cause: An attempt was made to choose a file for the analysis on this port
which has been selected for another port.
Action: Choose a different file.
Cause: The analyzer application is in the process of initializing the
instrument and is unable to locate the calibration files.
Action A: Click OK. Select Unit [n] > Calibration > Load from File and select
a file containing calibration data.
Action B: Click OK. Close the application and use the Setup program to
reinstall calibration files.
Cause: While attempting to zero the pressure transducers, the instrument was
unable to evacuate to a pressure of less than 1 mmHg. This may be
due to a leak or a bad calibration.
Action A: Check the sample tube fitting and ensure that it is securely attached to
the port.
Action B: Use the Setup program to reinstall the calibration files.