6041- Servo Calibration failed.
6042- Master pressure transducer calibration failed. Offset is out of range.
6043- Port 1 pressure transducer calibration failed. Offset is out of range.
6044- Port 2 pressure transducer calibration failed. Offset is out of range.
6045- Port 3 pressure transducer calibration failed. Offset is out of range.
6046- Po pressure transducer calibration failed. Offset is out of range.
6047- 10-torr pressure transducer calibration failed. Offset is out of range.
6048- Master pressure transducer calibration failed. Scale is out of range.
6049- Port 1 pressure transducer calibration failed. Scale is out of range.
6050- Port 2 pressure transducer calibration failed. Scale is out of range.
6051- Port 3 pressure transducer calibration failed. Scale is out of range.
6052- Po pressure transducer calibration failed. Scale is out of range.
6053- 10-torr pressure transducer failed. Scale is out of range.
6557- File [file name] already selected for the analysis
6558- Gas [n] in sample file [file name] does not match any gas in the unit.
Cause: There is a problem with the analyzer’s calibration.
Action: Use the Setup program to reinstall the calibration files. Contact a
Micromeritics service representative if this error message continues.
Cause: The same sample file is already assigned to a different port for this
analysis.
Action: Select a different sample file.
Cause: The analysis gas specified in the sample information file does not
match the analysis gas entered in the Unit Configuration.
Action: If the wrong adsorptive was selected in the sample information file,
change the adsorptive in the file.