SHOW SMART VACPREP STATUS
Smart VacPrep > Unit [n] > Show Status
The Smart VacPrep Status window allows the monitoring of degas operations for a manual degas
operation.
Field or Button Description
Cancel Cancels the current degas operation.
Check Click to check the outgassing rate of the samples on the selected port.
The following process occurs:
l The current degassing step is suspended.
l The vacuum valves are closed and the vacuum level on the selected
ports is monitored.
l The Status window indicates that the degassing operation is being
checked and displays the outgassing rate as it becomes available.
During the degas check, the Check button changes to Continue. When
Continue is clicked, the valves open and the degassing operation
resumes.
If the outgassing rate indicated that the sample has been freed of
contaminants (minimal pressure increases), click Skip to advance to the
next state of the degassing operation. For example, if degassing is
checked after the setpoint is attained, Skip advances you to the ramping
stage.
Sample The name of the sample file.
Skip Skips the current degas operation.
Status Indicates the current status of the degas.
Smart VacPrep Status Fields and Buttons Table
Show Smart VacPrep Status
Smart VacPrep Operator Manual 067-42800-01
Dec 2013
3 - 11