1. Turn off the power to the pump and disconnect the pump from the power source.
2. Loosen the clamp wing nut at the vacuum pump exhaust port. Swing the clamp away from the
exhaust port and remove it.
3. Remove and discard the exhaust filter. Ensure the centering ring remains in place on the exhaust
port.
4. Place the new filter on the centering ring.
5. Open the clamp and place it around the exhaust port flange and the exhaust filter flange. Swing
the clamp fastening screw toward the exhaust port until it fits into the slot in the other half of the
clamp. Tighten the wing nut securely by hand.
6. Turn on the power to the pump.
Replace the Vacuum Pump Exhaust Filter
Smart VacPrep Operator Manual — 067-42800-01
Dec 2013
6 - 11