BeneVision N1 Patient Monitor Operator’s Manual 13 - 11
Cannot see systolic pressure and diastolic
pressure for P1/P2/P3/P4
Set Measure to All in the P1/P2/P3/P4 setup page. For more information, see
13.5.3Setting the Pressure Type for Display.
IBP readings seem unstable 1. Make sure there are no air bubbles in the transducer systems.
2. Check that the transducer is properly fixed.
3. Zero the transducer again.
4. Replace the transducer.
Zeroing of IBP channel(s) fails. 1. Ensure that the channels are open to air.
2. Perform zero calibration again.
Do not sway the IBP transducer and
tubing during zero calibration.
For more information, see 13.2.3Zeroing
the IBP transducer
.
3. If zero calibration still fails, replace the transducer.