C - 6 BeneVision N1 Patient Monitor Operator’s Manual
C.4 Temperature Default Settings
C.4.1 Temperature Default Settings
C.5 NIBP Default Settings
PR Alarm switch On
High limit Adult: 120
Pediatric: 160
Neonate: 200
Low limit Adult: 50
Pediatric: 75
Neonate: 100
Priority Med
Alarm Outputs Off
Alarm Source Auto
PR Source Auto
Display PR On
QRS Volume General, OR: 2
Other departments: 0
Item Default Setting
Item Default Setting
Txx (xx refers to
termperature
site)
Alarm switch On
High limit 38.0 °C
Low limit 35.0 °C
Priority Med
Alarm Outputs Off
ΔT Alarm switch On
High limit 2.0 °C
Priority Med
Alarm Outputs Off
Item Default Setting
NIBP-S Alarm switch On
High limit Adult: 160 mmHg
Pediatric: 120 mmHg
Neonate: 90 mmHg
Low limit Adult: 90 mmHg
Pediatric: 70 mmHg
Neonate: 40 mmHg
Priority Med
Alarm Outputs Off