BeneVision N1 Patient Monitor Operator’s Manual C - 1
C Default Settings
C.1 ECG, Arrhythmia, ST and QT Default Settings
C.1.1 ECG Default Settings
Item Default Setting
HR Alarm switch On
High limit Adult: 120 bpm
Pediatric: 160 bpm
Neonate: 200 bpm
Low limit Adult: 50 bpm
Pediatric: 75 bpm
Neonate: 100 bpm
Priority Med
Alarm Outputs Off
Alarm Source Auto
ECG1 II
ECG2 (5-lead, 6-lead, 12-lead) V, Va, V1
Va (for 6-lead only) Va
Vb(for 6-lead only) Vb
ECG Gain ×1
Speed 25 mm/sec
Filter OR: Surgery
CCU: Diagnostic
Other departments: Monitor
Notch Filter On
Lead Set Auto
Smart Lead On
QRS Volume General and OR: 2
Other department : 0
Baseline Drift Removal (for 12-lead only) On
Waveform Layout (for 12-lead only) Standard
QRS Threshold 0.16 mV
Paced Adult: Unspecified
Pediatric/neonate: No
Pacer Reject Off