BeneVision N1 Patient Monitor Operator’s Manual C - 3
C.1.2.2 Arrhythmia Threshold Default Settings
C.1.3 ST Default Settings
Item
Default Setting
Adult Pediatric
Asystole Delay 5 sec 5 sec
Tachy 120 bpm 160 bpm
Brady 50 bpm 75 bpm
Extreme Tachy 160 bpm 180 bpm
Extreme Brady 35 bpm 50 bpm
Multiform PVC Window 15 beats 15 beats
PVCs/min 10 10
Pauses/min 8 8
Pause Threshold 2.0 sec 2.0 sec
AF/Irr Rhy End Time 2 min 2 min
V-Tach Rate 130 bpm 130 bpm
V-Brady Rate 40 bpm 40 bpm
V-Tach PVCs 6 6
V-Brady PVCs 5 5
Item Default Setting
ST Alarm Mode Absolute
ST Alarm Mode set to
Absolute:
ST-I, ST-II, ST-III, ST-aVR, ST-aVL, ST-aVF, ST-V1, ST-
V2, ST-V3, ST-V4, ST-V5,ST-V6, ST-Va, ST-Vb
Alarm switch Off
High limit 0.2 mV
Low limit -0.2 mV
Priority Med
Alarm Outputs Off
ST Alarm Mode set to
Relative:
ST Single,
ST Dual
Alarm switch Off
High limit 0.1 mV
Low limit -0.1 mV
Priority Med
Alarm Outputs Off
ST Analysis Off
ST Segment Auto
Show Marker Off
ST Point J+60
Auto Adjust On
J48
ISO -80

Table of Contents

Other manuals for Mindray BeneVision N1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N1 and is the answer not in the manual?

Mindray BeneVision N1 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N1
CategoryMedical Equipment
LanguageEnglish

Related product manuals