BeneVision N1 Patient Monitor Operator’s Manual 21 - 9
21.3.7.2 Setting the ST Alarm Delay Time
1. Select the Main Menu quick key → from the System column select Maintenance → input the required
password
→ select .
2. Select the
Alarm tab → Other tab.
3. Set
ST Alarm Delay.
21.3.8 Adjusting the Alarm Light Brightness
To adjust the alarm light brightness, follow this procedure:
1. Select the
Main Menu quick key → from the System column select Maintenance → input the required
password
→ select .
2. Select the
Alarm tab → Other tab.
3. From the
Alarm Light Brightness area, set Primary Screen (note that you may have to scroll up the dialog
to see this option). If you set alarm light brightness to
Auto, the monitor automatically adjusts the alarm
light brightness according to the ambient light. The stronger the ambient light is, the brighter the alarm
light is.
21.3.9 Changing Other Alarm Settings
21.3.9.1 Setting the Lethal Arrhythmia Alarms Switch
To allow a clinician to disable the lethal arrhythmia alarms, follow this procedure:
1. Select the
Main Menu quick key → from the System column select Maintenance → input the required
password
→ select .
2. Select the
Alarm tab→ select the Other tab.
3. From the
Other block, set Lethal Arrh Alarms Off as follows.
â—† Lethal Arrh Alarms Off is set to Disable by default. In this case, a clinician cannot switch off lethal
arrhythmia alarms.
â—† If you set Lethal Arrh Alarms Off to enable, a clinician can switch off lethal arrhythmia alarms from
the
ECG dialog. For more information, see 7.6.4.1Changing Arrhythmia Alarm Settings.
• If any of lethal arrhythmia alarms are switched off, the ECG waveform area displays a message to
indicate that corresponding alarms are off.
• If a clinician switches off all arrhythmia alarms, the monitor will not present arrhythmia alarms.
Always keep the patient under close surveillance.
21.3.9.2 Setting the Intubation Time
The default intubation time is 2 minutes. To change the intubation time, following this procedure:
1. Select the
Main Menu quick key → from the System column select Maintenance → input the required
password
→ select .
2. Select the
Alarm tab → Other tab.
3. Set the
Intubation Mode Period.
21.3.9.3 Setting Arrhythmia Alarm Timeout Period
The arrhythmia algorithm can disable alarm light and alarm tone for designated period of time when certain
arrhythmia alarms are detected.
To set the arrhythmia alarm timeout period, follow this procedure: