14 - 2 BeneVision N1 Patient Monitor Operator’s Manual
14.3 CO
2
Safety Information
WARNING
Route all tubing away from the patient’s throat to avoid strangulation.
CAUTION
Remove the airway sample line from the patient’s airway while nebulized medications are being
delivered. Measuring CO
2
during nebulization may lead to inaccurate CO
2
readings.
EtCO
2
values measured from the CO
2
module may differ from those from the blood gas analysis.
Avoid mechanical shock to the sidestream CO
2
module.
NOTE
The CO
2
module automatically suppresses physiological alarms until breathing waves have been
detected. Ensure that the patient is properly connected when using the CO
2
module.
14.4 CO
2
Measurement Limitations
The following factors may influence the measurement accuracy:
Leaks or internal venting of sampled gas
Mechanical shock
Cyclic pressure up to 10 kPa (100 cmH
2
O)
Other sources of interference, if any
For more information, refer to
A.14.8CO
2 Specifications
.
CAUTION
Measurement accuracy of the sidestream CO
2
module may be affected by the breath rate and
inspiration/expiration (I/E) ratio.
Measurement accuracy of the microstream CO
2
module may be affected by the breath rate.
14.5 CO
2
Display
The CO
2
numeric and waveform areas provide FiCO
2
measurement, EtCO
2
measurement, awRR measurement,
and a CO
2
waveform.
(1) CO
2
dialog hard key (2) CO
2
Measure/standby hard key
(3) CO
2
watertrap seat (4) Gas outlet
(5) Sample line connector (6) Internal CO
2
adapter connector
(1)
(4)
(2)
(3)