BeneVision N1 Patient Monitor Operator’s Manual 5 - 3
WARNING
Do not discharge a patient on the bedside monitor before patient information is completely
transferred. Discharging the patient during transfer will result in incomplete patient information.
NOTE
The system automatically switches on the HR/PR (high, low) alarms and lethal arrhythmia alarms
after transferring the patient data.
5.4 Exporting Patient Data
To export the data of the current patient and discharged patients, follow this procedure:
1. Connect the N1 to the Dock.
2. Connect the USB drive to the Dock’s USB connector.
3. Access the Discharged Patients dialog box by selecting the Main Menu quick key from the Patient
Management column select Discharged Patients.
4. From the patient list select desired patients.
5. Select Export Data.
5.5 Deleting Patient Data
To delete the data of discharged patients, follow this procedure:
1. Access the Discharged Patients dialog box by selecting the Main Menu quick key from the Patient
Management column select Discharged Patients.
2. From the patient list select desired patients.
3. Select Delete.
5.6 Stopping a Parameter Measurement
To stop monitoring a parameter, follow this procedure:
1. Remove corresponding sensor from the patient.
2. Disconnect the sensor from the patient cable.
3. Disconnect the patient cable from the parameter module.
5.7 Discharging a Patient
Before monitoring a new patient, discharge the previous patient. The technical alarms are reset, and monitor
settings return to their defaults. For more information, see 20.4Setting Default Configuration.
After a patient is discharged, the monitor automatically admits a new patient.
WARNING
Always discharge the previous patient before starting to monitor a new patient. Failure to do so can
lead to data being associated with the wrong patient.
To manually discharge a patient, use any of the following methods to access the Discharge Patient dialog:
Swipe down the touchscreen with two fingers.
Select the Discharge Patient quick key (only available when the N1is used with the independent external
display).
Select the patient information area at the top left corner of the screen Discharge Patient.