BeneVision N1 Patient Monitor Operator’s Manual 20 - 3
4. Select Export Configuration.
5. Select the configurations and
User Maintenance Settings to export.
6. Select
Export.
20.7.2 Importing a Configuration
To import the configuration from the USB drive to the monitor, follow this procedure:
1. Connect the N1 to the Dock.
2. Connect the USB drive to the Dock’s USB port.
3. Select the
Main Menu quick key → from the Configuration column select Manage → input the required
password
→ Select .
4. Select
Import Configuration.
5. Select the configurations and
User Maintenance Settings to import.
6. Select
Import.
20.7.3 Loading a Configuration
You may make changes to some settings during operation. However, these changes or the pre-selected
configuration may not be appropriate for the newly admitted patient. Therefore, the monitor allows you to load
a desired configuration to ensure that all the settings are appropriate for your patient.
To load a configuration, follow this procedure:
1. Select the
Main Menu quick key → from the Configuration column select Load.
2. Select the desired configuration.
◆ Select the configuration on this monitor in the Local page.
◆ Select the configuration on the USB drive in the USB Drive page.
3. Select
Load.
• The monitor may configure some settings by default when you load a configuration of different
software version with the current configuration.
20.8 Printing Configurations
To print both factory configurations and user configurations, follow this procedure:
1. Select the
Main Menu quick key →from the Configuration column select Manage → input the required
password
→ select .
2. Select
Print Configuration.
3. Select desired configurations.
4. Select
Print.
20.9 Modifying Configuration Password
To modify the configuration password, follow this procedure:
1. Select the
Main Menu quick key → from the Configuration column select Manage → input the required
password
→ Select .
2. Select
Modify Password.
3. Respectively input the old password and new password.
4. Select
OK.