21 - 8 BeneVision N1 Patient Monitor Operator’s Manual
2. Select the Alarm tab → Other tab.
3. Set
CMS/eGW Disconnected.
21.3.6.6 Setting the Switch of the SpO
2
Desat Alarm Off
You can choose whether switching off the SpO
2
Desat alarm is permissible or not. To do so, follow this
procedure:
1. Select the M
ain Menu quick key → from the System column select Maintenance → input the required
password → select .
2. Select the Alarm tab → select the Other tab.
3. Set SpO2 Desat Alarm Off.
â—† Disable: the
SpO
2
Desat alarm is always on. The clinician cannot switch it off.
â—† E
nable: the clinician can switch off the SpO
2
Desat alarm. For more information, see
10.6.1Changing the SpO
2
Alarm Settings.
• If a clinician switches off the SpO2 Desat alarm, the monitor will not alarm when the patient’s SpO
2
is
extremely low. This may result in a hazard to the patient.
21.3.6.7 Setting the Switch of the Zero RR Alarm Off
You can choose whether switching off the zero RR alarm is permissible or not. To do so, follow this procedure:
1. Select the
Main Menu quick key → from the System column select Maintenance → input the required
password
→ select .
2. Select the
Alarm tab → select the Other tab.
3. Set
Zero RR Alarm Off.
â—† Disable: the zero RR alarm is always on. The clinician cannot switch it off.
◆ Enable: the clinician can switch off the zero RR alarm. For more information, see 9.5.1Setting the Resp
Alarm Properties
.
• If a clinician switches off the zero respiration rate alarm, the monitor will not issue the zero
respiration rate alarm when that situation occurs. This may result in a hazard to the patient. Keep the
patient under close surveillance.
21.3.7 Changing Alarm Delay Settings
21.3.7.1 Setting the Alarm Delay Time
For continuously measured parameters, you can set the alarm delay time. If the alarm condition is resolved
within the delay time, the monitor does not present the alarm.
To set the alarm delay time, follow this procedure:
1. Select the
Main Menu quick key → from the System column select Maintenance → input the required
password
→ select .
2. Select the
Alarm tab → Other tab.
3. Set
Alarm Delay.
The setting of
Alarm Delay is not applied to the zero respiration rate alarms and the ST alarms. You can set Zero
RR Delay and ST Alarm Delay separately.