22 - 4 BeneVision N1 Patient Monitor Operator’s Manual
22.7 Storing the Battery
When storing the battery, make sure that the battery terminals do not come into contact with metallic objects. If
the battery are stored for an extended period of time, place the battery in a cool place with a partial charge of
40% to 60% capacity.
Condition the stored battery every three months. For more information, see
22.6.1 Conditioning the Battery.
NOTE
• Remove the battery from the equipment if the equipment is not used for a prolonged time (for
example, several weeks). Otherwise the battery may overdischarge.
• Storing the battery at high temperature for an extended period of time will significantly shorten
their life expectancy.
• Storing the battery in a cool place can slow the aging process. Ideally the battery should be stored at
15 °C.
22.8 Recycling the Battery
Discard a battery in the following situations:
â–  The battery has visual signs of damage.
â–  The battery fails.
â–  The battery is aged and its runtime significantly less than the specification.
â–  The battery has been used for more than it service time.
Properly dispose of the battery according to local regulations.
WARNING
• Do not open the battery, heat the battery above 60 °C, incinerate the battery, or short the battery
terminals. They may ignite, explode, leak or heat up, causing personal injury.

Table of Contents

Other manuals for Mindray BeneVision N1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BeneVision N1 and is the answer not in the manual?

Mindray BeneVision N1 Specifications

General IconGeneral
BrandMindray
ModelBeneVision N1
CategoryMedical Equipment
LanguageEnglish

Related product manuals