14 - 4 BeneVision N1 Patient Monitor Operator’s Manual
3. Connect the other end of the gas sample line to the patient.
◆ For intubated patients requiring an airway adapter, install the airway adapter between the patient
circuit and the ventilator Y-piece.
◆ For non-intubated patients, place the nasal cannula onto the patient.
4. Connect the gas outlet to the scavenging system using an exhaust tube.
After the CO
2
module is connected, it enters measure mode by default and the monitor displays CO2 Starting.
CO
2
can be measured after the start-up sequence is complete.
• Do not use the adult or pediatric watertrap with a neonate patient. Otherwise, patient injury could
result.
• Connect the gas outlet to the scavenging system when measuring CO
2
using the sidestream CO
2
module.
• Check the compatibility of the CO
2
adapter and the sampling line before use. The CO
2
adapter is
intended for connecting an Oridion CO
2
sampling line.
(1) CO
2
adapter (2) Gas sample line
(1) Sample line (2) Connection to the ventilator
(3) Airway adapter (4) Connection to the patient