14 - 4 BeneVision N1 Patient Monitor Operator’s Manual
3. Connect the other end of the gas sample line to the patient.
For intubated patients requiring an airway adapter, install the airway adapter between the patient
circuit and the ventilator Y-piece.
For non-intubated patients, place the nasal cannula onto the patient.
4. Connect the gas outlet to the scavenging system using an exhaust tube.
After the CO
2
module is connected, it enters measure mode by default and the monitor displays CO2 Starting.
CO
2
can be measured after the start-up sequence is complete.
WARNING
Do not use the adult or pediatric watertrap with a neonate patient. Otherwise, patient injury could
result.
Connect the gas outlet to the scavenging system when measuring CO
2
using the sidestream CO
2
module.
CAUTION
Check the compatibility of the CO
2
adapter and the sampling line before use. The CO
2
adapter is
intended for connecting an Oridion CO
2
sampling line.
(1) CO
2
adapter (2) Gas sample line
(1) Sample line (2) Connection to the ventilator
(3) Airway adapter (4) Connection to the patient
(1)
(2)
(1)
(2)
(3)
(4)