BeneVision N1 Patient Monitor Operator’s Manual C - 9
IBP-D Alarm switch On
High limit • Art/Ao/UAP/BAP/FAP/LV/P1-P4 arterial pressure
Adult: 90 mmHg
Pediatric: 70 mmHg
Neonate: 60 mmHg
•PA
Adult: 16 mmHg
Pediatric and neonate: 4 mmHg
Low limit • Art/Ao/UAP/BAP/FAP/LV/P1-P4 arterial pressure
Adult: 50 mmHg
Pediatric: 40 mmHg
Neonate: 20 mmHg
•PA
Adult: 0 mmHg
Pediatric and neonate: -4 mmHg
Priority Med
Alarm Outputs Off
IBP-M Alarm switch On
High limit • Art/Ao/UAP/BAP/FAP/LV/P1-P4 arterial pressure
Adult: 110 mmHg
Pediatric: 90 mmHg
Neonate: 70 mmHg
•PA
Adult: 20 mmHg
Pediatric and neonate: 26 mmHg
•CVP
Adult: 14 cmH
2
O
Pediatric and neonate: 5 cmH
2
O
• ICP/RAP/LAP/UVP/P1-P4 venous pressure
Adult: 10 mmHg
Pediatric and neonate: 4 mmHg
Low limit • Art/Ao/UAP/BAP/FAP/LV/P1-P4 arterial pressure
Adult: 70 mmHg
Pediatric: 50 mmHg
Neonate: 35 mmHg
•PA
Adult: 0 mmHg
Pediatric and neonate: 12 mmHg
•CVP
Adult: 0 cmH
2
O
Pediatric and neonate: 0 cmH
2
O
• ICP/RAP/LAP/UVP/P1-P4 venous pressure
Adult: 0 mmHg
Pediatric and neonate: 0 mmHg
Priority Med
Alarm Outputs Off
Item Default Setting