11-3
3. Plug the mini water trap to the CO
2
sampling line connector on the monitor.
4. Connect the other end of the sampling line to the patient.
5. Connect the CO
2
gas outlet to the scavenging system with an exhaust tube.
CO
2
can be measured after the start-up is complete.
CAUTION
The DRYLINE PRIME mini water trap is automatically blocked when its service
life is reached. Replace the DRYLINE PRIME mini water trap if the alarm ‘‘CO2:
Change Watertrap’’ occurs.
The water trap has a filter preventing bacterium, water and secretions from
entering the module. Extended use could destroy the filter in water trap and fail
to stop the bacterium, water and secretions entering the module, result in
damaging the gas module and having infection risk.
Sampling line
Mini water trap