15-1 
15 Battery 
 
 
15.1 Overview 
The monitor is designed to operate from battery power when AC power supply is not 
available. In case of power failure, the equipment will automatically run power from 
internal battery. So we recommend you always install a fully charged battery in the 
equipment. 
 
Where the integrity of the external protective conductor in the installation or its 
arrangement is in doubt, the equipment shall be operated from the battery.   
 
 CAUTION 
  Remove the battery before transporting the equipment or if the equipment will 
not be used for a long time. 
 
NOTE 
  It is recommended to always install a fully charged battery in the monitor to 
ensure normal monitoring in case of accidental power failure. 
 
The on-screen battery symbol indicate the battery status as follows: 
    Indicates that battery works correctly. The solid portion represents the 
current charge level of the battery in proportion to its maximum charge level. 
    Indicates that the battery has low charge level and needs to be charged. 
    Indicates that the battery is almost depleted and need to be charged 
immediately. Otherwise, the monitor will shut down automatically. 
    Indicates that no battery is installed. 
 
The capacity of the battery is limited. When the battery is low, the technical alarm area 
displays [Low Battery], the alarm lamp flashes, and monitor produces alarm sound. 
 
If the battery is deplete, the battery symbol on the screen starts to flash, the technical 
alarm area displays [Battery Depleted], the alarm lamp flashes, and monitor produces 
alarm sound. At this moment, connect the equipment to AC mains to run the equipment 
and charge the battery. Otherwise the equipment will shut down.