14-4
2. Select [Network>>][NTP Setup>>].
3. Select [NTP Setup>>] to enter the NTP Setup menu. Set the following properties:
Set IP address and port.
Set the interval.
Set the time zone.
Switch on/off the daylight saving time.
4. Select [Ok] to exit the menu.
To manually start one time synchronization with the NTP server, follow this procedure:
1. Select [Main][Maintenance >>][User Settings >>]Enter the required
passwordSelect [Ok] to access User Settings menu.
2. Select [Network>>][NTP Setup>>][Manual].
14.3.4 Realtime Date Send Interval (for HL7 only)
The monitor will send out the monitored parameter data at the set intervals.
1. Select [Main][Maintenance >>][User Settings >>]Enter the required
passwordSelect [Ok] to access User Settings Menu.
2. Select [Network>>][EMR Communication>>].
3. Set [Real Data Send] to an appropriate setting.
14.3.5 Alarm Server Setup (for HL7 only)
1. Select [Main][Maintenance >>][User Settings >>]Enter the required
passwordSelect [Ok] to access User Settings menu.
2. Select [Network>>][Alarm Server Setup].
3. Set the IP address and port.
4. Select [Ok] to exit the menu.
14.4 Connecting the monitor to the CMS
To connect the monitor to the CMS, proceed as follows:
1 Select [Main][Maintenance >>][User Settings >>]enter the required
passwordSelect [Ok][Network >>][Monitor Network Setup >>].
2. Set [Network Type] and [Address Type].
3. Input the monitor IP address, subnet mask and gateway address if the [Address Type]
is set to [Manual]
4. Connect the monitor to the CMS through either of the following methods: