9-7
Patient Category Range (mmHg) Default (mmHg)
Adult 80 - 280 160
Pediatric 80 - 210 140
Neonate 60 - 140 90
NOTE
For known hypertensive patients, you need to set initial cuff pressure to a
higher value to reduce the measurement time.
9.7.3 Setting NIBP End Tone
The monitor can issue a reminder tone at the completion of NIBP measurement. The NIBP
End Tone is off by default. You can switch it on by accessing the [NIBP Setup] menu.
9.7.4 Switching On/Off Measurement on Clock
In auto measuring mode, if the clock is enabled, the NIBP automatic measurement
interval will be synchronized with the real time clock.
For example, if [Clock] is set to [On], and [Interval] is set to [20min], and then you start an
NIBP auto measurement at 14: 03, the next measurement will be taken at 14: 20, and the
following measurement time will be 14:40, 15:00 and so on.
9.7.5 Configuring a Custom Program
In [NIBP Setup] menu, you can select [Custom Program>>] and configure the duration
of automatic measurement cycle, and the time interval between two NIBP measurement.
You can define two programs, respectively program 1 and program 2. Each program can
at most include five cycles: A, B, C, D, and E. In each cycle, the [Duration] and [Interval]
can be set individually.
You shall start the programmed NIBP measurement manually, and then the monitor will
automatically perform the measurement based on the cycle and interval you have
defined.