Measurement accuracy may be affected by the breath rate and inspiration/expiration 
ratio (I: E ratio) as follow: 
 
EtCO
2
 value is within specification for breath rate ≤ 60 bpm and I/E ratio ≤ 1:1; 
  EtCO
2
 value is within specification for breath rate ≤ 30 bpm and I/E ratio ≤ 2:1. 
 
Measurement accuracy is unspecified for breath rate larger than 60 bpm. 
 
11.4 CO
2
 Display 
The CO
2
 parameter and waveform areas provide FiCO
2 
measurement, EtCO
2
 measurement, 
awRR measurement, and a CO
2
 waveform. 
 
 
 
 
 
(1)  End tidal CO
2 
(EtCO
2
) 
(2)  Fraction of inspired CO
2
 (FiCO
2
) 
(3)  Airway respiration rate (RR) 
(4) CO
2
 waveform 
(5) Scale  
 
 
11.5 Measuring CO
2
 
 
CAUTION 
  Eliminate the exhausted gas before performing the measurement. 
  Connect an exhaust tube to the gas outlet connector of the monitor to vent the 
calibration gases to a scavenging system. 
  Check that the alarm limit settings are appropriate before taking measurement. 
 
To measure CO
2
, follow this procedure: 
1.  Select the appropriate sampling line according to the patient category. 
2.  Connect the sampling line to the gas sampling port of the mini water trap by 
clockwise turning.     
(1)
2
4
5 
(3)