5-4
5.7 Restoring the Latest Configuration Automatically
During operation, you may make changes to some settings. However, these changes may
not be saved as user configuration. To prevent the changes from losing in case of a
sudden power failure, the equipment stores the configuration in real time. The saved
configuration is the latest configuration.
The monitor restore the latest configuration if restarts within 60 seconds after the power
failure. And it will restore the default configuration rather than the latest configuration if
restarts 120 seconds later after the power failure. The monitor may load either the latest
configuration or the default configuration if restarts from 60-120 seconds after the power
failure.