E-3
E.5 Earth Leakage Test
Run an Earth Leakage test on the device being tested before performing any other
leakage tests.
The following outlet conditions apply when performing the Earth Leakage test:
normal polarity (Normal Condition),
reverse polarity (Normal Condition),
normal polarity with open neutral (Single Fault Condition),
reverse polarity with open neutral (Single Fault Condition)
LIMITS
For UL60601-1,
300 A in Normal Condition
1000 A in Single Fault Condition
For IEC60601-1,
500 A in Normal Condition
1000 A in Single Fault Condition
E.6 Patient Leakage Current
Patient leakage currents are measured between a selected applied part and mains earth.
All measurements have a true RMS only
The following outlet conditions apply when performing the Patient Leakage Current test.
normal polarity (Normal Condition);
reverse polarity (Normal Condition),
normal polarity with open neutral (Single Fault Condition);
reverse polarity with open neutral (Single Fault Condition).
normal polarity with open earth (Single Fault Condition);
reverse polarity with open earth (Single Fault Condition).
LIMITS
For CF applied parts
10A in Normal Condition
50A in Single Fault Condition