E.5 Earth Leakage Test 
Run an Earth Leakage test on the device being tested before performing any other 
leakage tests. 
 
The following outlet conditions apply when performing the Earth Leakage test:   
  normal polarity (Normal Condition), 
  reverse polarity (Normal Condition), 
  normal polarity with open neutral (Single Fault Condition), 
  reverse polarity with open neutral (Single Fault Condition) 
 
LIMITS 
For UL60601-1, 
  300 A in Normal Condition 
  1000 A in Single Fault Condition 
 
For IEC60601-1, 
  500 A in Normal Condition 
  1000 A in Single Fault Condition 
 
 
E.6 Patient Leakage Current 
Patient leakage currents are measured between a selected applied part and mains earth. 
All measurements have a true RMS only   
 
The following outlet conditions apply when performing the Patient Leakage Current test.   
  normal polarity (Normal Condition); 
  reverse polarity (Normal Condition), 
  normal polarity with open neutral (Single Fault Condition);   
  reverse polarity with open neutral (Single Fault Condition). 
  normal polarity with open earth (Single Fault Condition);   
  reverse polarity with open earth (Single Fault Condition). 
 
LIMITS 
For CF    applied parts 
  10A in Normal Condition   
  50A in Single Fault Condition